EGA-1012 / EGPS-1012电子枪

Electron Gun System: 5 eV to 1000 eV

Quick Overview

High current with medium diameter spot size

Energy Range: 5 eV to 1000 eV
Beam Current: 1 μA to 2 mA
Spot size: 10 mm to 25 mm
Mounting: 2¾CF
Features/Options: Flood beam, Pulsing, Dual Grid

 

EGA-1012 ELECTRON GUN SPECIFICATIONS:
BEAM ENERGY :5 eV to 1000 eV (Independently adjustable)
BEAM CURRENT: 1 μA to 2 mA (Independently adjustable)
ENERGY SPREAD: Approx. cathode thermal spread, calculated: Y2O3 – 0.4eV
SPOT SIZE :10 mm to 25 mm at 55 mm working distance
WORKING DISTANCE:10 mm to 200 mm
PULSE CAPABILITY (using appropriate pulse generator, not included) 
Optional Dual grid power supply: pulse width 2 µs to DC, rise / fall 5—ns, rep rates to 5 kHz (TTL required). 
BEAM DEFLECTION :None
FIRING UNIT :Customer-replaceable Firing Unit Cartridge includes precision-aligned cathode, Wehnelt (G-1) assembly and anode
CATHODE TYPES :Standard: Yttrium Oxide
MOUNTING :Standard: 2¾ inch CF flange 
Optional: Unmounted
FARADAY CUP:Various stand-alone designs available
INSERTION LENGTH :29 mm
BEAM ALIGNMENT: Optional mechanical alignment with ± 2° Port Aligner
GUN DIAMETER:33.5 mm at gun end
FEEDTHROUGHS: Multi-pin brazed ceramic, threaded stainless steel shell
CABLES /CONNECTORS:
Multi-conductor high voltage fully ground-shielded, with mating aluminum connector to connect gun and power supply. Standard lengths: 3 m, Optional: 5 m
MAXIMUM BAKEOUT :350°C with cables removed (Removable Faraday cup pneumatic actuator 65°C)

 

EGPS-1012 ELECTRON GUN POWER SUPPLY SPECIFICATIONS:
OUTPUT :All necessary voltages to drive the EGA-1012 Electron Gun
ENERGY STABILITY: ±0.01% per hour ±0.02% per 8 hours at full output
BEAM STABILITY: ±1.0% per hour with Emission Current Control or ±10% per hour after warm up without ECC
CONTROLS: Energy, Source, Grid, Emission Current Control (ECC)
METERING :FlexPanel digital meters: Energy, Source Voltage, Source Current, Emission Current, Grid Voltage
COMPUTER/REMOTE CONTROL
Power supplies: 0 to +10 V 
Metering: 0 to +2 V 
Standard 50-pin connector for analog input/output and RS-232 serial port (RS-422 or RS-485 available, if specified at time of order)
INPUT :115 VAC or 230 VAC, 50 to 60 Hz single phase, 250 VA
ENVIRONMENT :Temperature: 0 to 40ºC, Relative humidity: 0 to 75% RH non condensing 
Classified as a pollution degree 2, installation category (overvoltage category) II environment unit
SOFTWARE :Standard configuration designed for RS-232 connections. 
Optional: National Instruments LabVIEWTM file, designed to run with computer DAQ boards NI PCI-6713 and PCIe-6341. SCSI interface
DIMENSIONS (width x height x depth):
17 in. x 3.5 in. x 22 in.excluding handles (425 mm x 90 mm x 560 mm); 19 in. rack mountable

Related posts